Technical Specifications
Data transmission
- Multi instrument networking capability (up to 40 apparatus) via a Radio-link or a RS485 cable;
- Data transmitted by radio frequency can travel through walls and floors, over on long distances depending on the environment. For example, 900 MHz signals will travel from 250 m when traveling across 2 concrete walls, up to 1 km in open air. This span can be increased considerably with the use of an RF repeater;
Operation
- Each probe is easily calibrated by one or two points;
- Menu-driven programming environment;
- FLASH memory for stocking data (49 152 data for 25 days);
- Dissolved oxygen measurements are computed in both mg/L and % saturation;
- temperature in oC or oF, salinity in PSU and flow in L/min. or US Gal/min;
- The Monitron is easy to understand and to operate. A detailed instruction manual is provided. Service is also available for commissioning and/or training if the need arises.
Advanced features
- Enviro-Monitron’s advanced control functions include PID (proportional derivative integral control), cyclic PID (voltage variable) and on/off regulation.
- These sophisticated features integrate the advantage of great precision and stability in maintaining DO, temperature, % of recirculation, salinity, etc., at a given value.
- This is done with extreme ease: the operator simply enters the set points for a given parameter; a smooth and precise response is then automatically set in motion.
Auxiliary I/O supporting 18 Inputs (8 Digitals, 8 Analogs and 2 Pulses);
Inputs can monitor parameters such as pump, chillers and heater’s status, water levels. |
| Every unit is built to client’s specification. Most popular configurations include DO, pH, % recirc, and temperature control. |
| 8 Power Switches with International rectifier; output 5 Amp. 24 Volts DC each; |
| Intergrated data monitoring (8 analog inputs, 8 digital inputs and 2 high-speed digital inputs); |
| process control (6 cyclic or proportional outputs, 2 analog 4-20 mA, and 2 on/off outputs), display, alarming, and the capacity to transfer data to a central computer via a radio link or cable; |
| The Enviro-Monitron will accept any probe that meets the input specifications of 0- 5 V and 4-20 mA; |
|